Injection-Valve
The two-position 6-port injection valve is driven by a stepper motor. Only five
of the six ports are used (port 3 is not used). A lever/slider mechanism
transfers the movement of the stepper motor to the injection valve. Two
microswitches monitor switching of the valve (bypass and mainpass end
positions).
No valve adjustments are required after replacing internal components.
Table 2
Injection-Valve Technical Data
Motor type
Seal material
Stator material
Number of ports
Switching time
Standard
4 V, 1.2 A stepper motor
Vespel™ (Tefzel™
available)
Ceramic/PEEK
6
< 150 ms
Introduction to the Autosampler
Preparative MBB™
4 V, 1.2 A stepper motor
PEEK
PEEK
6
< 150 ms
Sampling Unit
Autosampler SL
4 V, 1.2 A stepper motor
PEEK
None
6
< 150 ms
1
15