Cardiac Science Powerheart G3 Plus 9390E Operator's And Service Manual page 79

Aed
Hide thumbs Also See for Powerheart G3 Plus 9390E:
Table of Contents

Advertisement

Table 14: Ultra-low Variable Energy (150 VE) Powerheart G3 Waveform
Patient's
Impedance
(Ohms)
25
50
75
100
125
150
175
Table 15: Low Variable Energy (200 VE) Powerheart G3 Waveform
Patient's
Impedance
(Ohms)
25
50
75
100
125
150
175
Version 112-2025-109 B
Phase 1
Voltage*
Duration*
(Volts)
(MS)
1393
3.3
1420
4.5
1430
5.8
1434
7.0
1437
8.3
1439
9.5
1441
10.8
Phase 1
Voltage*
Duration*
(Volts)
(MS)
1609
3.3
1640
4.5
1651
5.8
1656
7.0
1660
8.3
1662
9.5
1663
10.8
Phase 2
Voltage*
Duration*
(Volts)
(MS)
743
3.2
909
3.2
973
3.2
1007
3.2
1027
3.2
1040
3.2
1049
3.2
Phase 2
Voltage*
Duration*
(Volts)
(MS)
858
3.2
1050
3.2
1124
3.2
1163
3.2
1186
3.2
1201
3.2
1212
3.2
Energy**
(Joules)
145-196
128-173
116-156
108-146
102-138
98-132
95-128
Energy**
(Joules)
193-260
170-230
155-209
144-194
136-184
131-176
126-170
Technical Data
7-9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Powerheart g3 plus automatic 9390a

Table of Contents