Cardiac Science Powerheart G3 Plus 9390E Operator's And Service Manual page 77

Aed
Hide thumbs Also See for Powerheart G3 Plus 9390E:
Table of Contents

Advertisement

Table 13: Parameters (continued)
Parameter
Cardiac Rhythms Used to
Test the Rhythm
Recognition Detection
System for Powerheart G3
AEDs
Version 112-2025-109 B
Detail
Shockable Rhythm* – VF: Meets AAMI DF 39
requirement and AHA recommendation of
Sensitivity of >90%
(*Automatic External Defibrillators for Public
Access Defibrillation: Recommendations for
Specifying and Reporting Arrhythmia Analysis
Algorithm Performance, Incorporating New
Waveforms and Enhancing Safety, AHA AED Task
Force and approved by the AHA Science Advisory
and Coordinating Committee. Circulation,
1997(95), pp 1677-1682)
Shockable Rhythm – VT: Meets AAMI DF 39
requirement and AHA recommendation of
Sensitivity of >75%
Non-shockable Rhythm – NSRMeets AAMI DF 39
requirement (>95%) and AHA recommendation
(>99%) of Specificity
Non-shockable – Asystole: Meets AAMI DF 39
requirement and AHA recommendation of
Specificity of >95%
Non-shockable: Meets AAMI DF 39 requirement
and AHA recommendation of Specificity – all
other rhythms of >95%
For detailed information contact Cardiac Science
for white papers:
P/N 112-2013-001 (Pediatric Defibrillation
Instructions)
P/N 110-0033-001 (RHYTHMx White Paper)
P/N 400781 (STAR Biphasic White Paper)
Technical Data