Table 16: High Variable Energy (300 VE) Powerheart G3 Waveform
Patient's
Impedance
(Ohms)
25
50
75
100
125
150
175
* All values are typical.
**Allowable energy range.
7-10
Star Biphasic Waveform
Phase 1
Voltage*
Duration*
(Volts)
(MS)
1869
3.3
1906
4.5
1918
5.8
1925
7.0
1928
8.3
1931
9.5
1933
10.8
Phase 2
Voltage*
Duration*
(Volts)
(MS)
997
3.2
1220
3.2
1306
3.2
1351
3.2
1378
3.2
1396
3.2
1408
3.2
Version 112-2025-109 B
Energy**
(Joules)
260-351
230-311
210-283
195-263
184-248
176-238
170-230