Immunity test
Radiated RF
IEC 61000-4-3
Version 112-2025-109 B
IEC 60601 test
level
10 V/m
80 MHz to 2.5 GHz
Compliance level
10 V/m
Product Information and Safety 1-15
Electromagnetic environment—
guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
AED, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance
d = 1.2
P
80 MHz to 800 MHz
d = 2.3
P
800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m)
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
b
.
c
d