Table 5: Preparation (continued)
Voice Prompt
"Firmly Place the Pad
Without the Liner on the
Patient, Exactly as
Illustrated. This Pad Can
be Placed on Either of the
Two Locations Shown."
"Next, Peel the Blue
Plastic Liner Off of the
Second White Pad."
"Firmly Place the Second
Pad on the Opposite
Location, Exactly as
Illustrated."
Table 6: Analysis
Voice Prompt
"Do Not Touch Patient!
Analyzing Heart Rhythm.
Please Wait."
"Preparing Shock. Move
Away from the Patient!"
Table 7: Delivering Shock - Semi-Automatic
Voice Prompt
"Press Red Flashing Button
to Deliver Shock."
"Shock Delivered"
3-14
Voice Prompts and Text Display
Text Display
PRESS PAD FIRMLY
TO CHEST AS SHOWN
PEEL SECOND PAD OFF
BLUE PLASTIC LINER
PRESS PAD FIRMLY
AS SHOWN
Text Display
DO NOT TOUCH PATIENT
ANALYZING RHYTHM
NO CONTACT
WITH THE PATIENT
Text Display
PRESS BUTTON
TO DELIVER SHOCK
SHOCK DELIVERED
Situation
Prompts the rescuer to place one pad
on the patient.
Prompts the rescuer to remove the liner
from the second pad.
Repeats until second pad placement is
sensed. If the pad is placed before
prompt starts then this prompt will be
skipped. This prompt will be
interrupted when second pad is placed.
Situation
Repeats until analysis of the patient's
cardiac rhythm is completed. This
prompt will be interrupted when ready
to shock.
Repeats while the AED is preparing to
deliver a defibrillation shock (charging).
Situation
Prompts after the AED is fully charged
and ready to deliver the defibrillation
shock. The RED SHOCK indicator
flashes and the phrase repeats for 30
seconds or until the SHOCK button is
pushed.
Prompts when the shock is delivered.