Table 13: Parameters (continued)
Parameter
Environmental Operation
and Standby Conditions
Shipment and Transport
environmental Conditions
(for up to 1 week)
Pads
9146 Lithium Battery
Specifications
(p/n: 9146-001)
Batteries and Capacitor
Charge Times
7-2
Parameters
Detail
Temperature: 0°C to 50°C (32°F to 122°F)
Humidity: 5% to 95% (non-condensing)
Pressure: 57kPa (+15,000ft) to 103kPa (-500ft)
Temperature: -30°C to 65°C (-22°F to 149°F)
Humidity: 5% to 95% (non-condensing)
Pressure: 57kPa (+15,000ft) to 103kPa (-500ft)
Self-adhesive, disposable defibrillation pads
Minimum combined surface area: 228cm
Extended length of lead wire: 1.3m
Output voltage: 12VDC (max)
Batteries are non-rechargeable
Lithium contents: 9.2g (max)
Check local regulations for disposal information
Full Operational Replacement Guarantee (from
date of installation): 4 Years
Estimated Shelf Life (from date of manufacture): 5
Years
Typical Shocks: up to 290 shocks
The battery operating life depends on the
Note:
type of battery, actual usage and environmental
factors.
A new battery typically takes 10 seconds to
charge the AED to maximum energy.
A battery with reduced capacity causes the red
LED light to initially turn ON and typically takes
13 seconds to charge a fully discharged AED to
maximum energy.
2
Version 112-2025-109 B