Close the lid.
3
Confirm that the STATUS INDICATOR turns RED.
4
Open the lid and confirm that the Pad indicator is lit.
5
Reconnect the pads and close the lid.
6
Make sure the expiration date is visible through the clear window of the
7
lid.
Check to make sure that the STATUS INDICATOR is GREEN. If the
8
pads are not installed properly, the PAD indicator will illuminate; call
Customer Service for assistance.
Open the lid and confirm that no diagnostic indicators are lit.
9
Check the expiration date of the pads; if expired, replace them.
10
Check the pads packaging integrity.
11
Close the lid
12
Check the Integrity of the Service Indicator (LED) and Circuitry:
Immediately after opening the AED lid, press and hold the Shock button
1
and confirm that the Service LED is lit.
Release the Shock/Continue button.
2
Close the lid.
3
Verify that the STATUS INDICATOR remains RED.
4
Open the lid and confirm that no diagnostic indicators are lit.
5
Close the lid.
6
Verify that the STATUS INDICATOR turns GREEN.
7
Check the Integrity of the Case:
Examine the molded case of the AED for any visible signs of stress. If the
case shows signs of stress, contact Cardiac Science Customer Service
(See Contact Information on page 1-1) or contact your local Cardiac Science
distributor.
CAUTION: Case Cleaning Solutions.
!
When disinfecting the case, use a non-oxidizing disinfectant, such as ammonium
salts or a glutaraldehyde based cleaning solution, to avoid damage to the metal
connectors.
6-4
Scheduled Maintenance
Version 112-2025-109 B