The following table lists the voice and text prompts and a description of
when the prompts are issued.
Table 4: Initial Instructions
Voice Prompt
"Stay Calm. Follow These
Voice Instructions. Make
Sure 911 is Called Now!"
"Stay Calm. Follow These
Voice Instructions. Make
Sure Emergency Services
are Called Now!"
Table 5: Preparation
Voice Prompt
"Begin by Exposing
Patient's Bare Chest and
Torso. Remove or Cut
Clothing if Needed."
"When Patient's Chest and
Torso are Exposed,
Remove Square Foil
Package from Lid of AED."
"Tear Open Foil Package
Across Dotted Line and
Remove Pads."
"Next, Separate One of
the White Pads
Completely from Blue
Plastic Liner. Begin Peeling
from the Tabbed Corner."
Version 112-2025-109 B
Text Display
CALL 911!
CALL EMERGENCY
SERVICES NOW!
Text Display
BARE PATIENT'S TORSO
REMOVE CLOTHING
WHEN CHEST IS BARE
REMOVE FOIL PACKAGE
TEAR OPEN PACKAGE
REMOVE PADS
PEEL ONE PAD FROM
BLUE PLASTIC LINER
Situation
Plays after lid opening self test, default
ON.
Medical Director may use MDLink to
select this prompt instead of "CALL
911!".
MDLink also allows the emergency
services and 911prompt to be disabled.
Situation
Prompts the rescuer to remove patient
clothing.
Prompts the rescuer to remove the
pads from AED lid.
Prompts the rescuer to open the pad
package and remove pads.
Repeats every 3 seconds until the pads
are separated. If a pad has been peeled
before the prompt starts, this prompt
will be skipped. This prompt will be
interrupted when pad is peeled.