Table 10: Pad Issues
Voice Prompt
"Press Pads Firmly to
Patient's Bare Skin."
Table 11: Other Prompts
Voice Prompt
"Battery Low"
"Analysis Interrupted.
Stop Patient Motion."
"Open Lid to Continue
Rescue."
"Rhythm Changed.
Shock Cancelled."
"Remove Cable to
Continue Rescue."
Version 112-2025-109 B
Text Display
PRESS PADS FIRMLY
TO BARE SKIN
Text Display
BATTERY LOW
ANALYSIS INTERRUPTED
STOP PATIENT MOTION
OPEN LID
TO CONTINUE RESCUE
RHYTHM CHANGED
SHOCK CANCELLED
REMOVE CABLE
TO CONTINUE RESCUE
Situation
Prompts when better pad connectivity
to the patient's skin is required because
impedance is too high.
Situation
Occurs once when the battery voltage
becomes low, although a rescue can
continue for approximately 9 more
shocks. When the battery is too low to
do a rescue, the following will occur:
1 "BATTERY LOW" will show on the LCD
STATUS INDICATOR will turn RED
2
AED will BEEP once every 30
3
seconds
You must replace the battery before
continuing with the rescue. If
completely depleted, all AED activity
will terminate.
When the AED detects ECG noise
artifact, stop moving or touching the
patient. Remove other electronic
devices within a 5 meter radius.
When the lid is inadvertently closed
during a rescue, this prompt will repeat
for 15 seconds.
When the device is prepared to shock
then detects a change in rhythm and
therefore cancels the shock.
When a serial communication cable is
connected to the AED during a rescue,
the phrase repeats until the cable is
disconnected.
Getting Started 3-17