WARNING: Electric Shock and Fire Hazard.
!
Do not connect any telephones or unauthorized connectors to the socket on this
equipment.
WARNING: Battery is Not Rechargeable.
!
Do not attempt to recharge the battery. Any attempt to recharge the battery may
result in an explosion or fire hazard.
WARNING: Shock Hazard.
!
Do not disassemble the AED. Failure to observe this warning can result in
personal injury or death. Refer maintenance issues to Cardiac Science authorized
service personnel.
WARNING: Possible Radio Frequency (RF) Susceptibility.
!
RF susceptibility from cellular phones, CB radios, and FM 2-way radio may cause
incorrect rhythm recognition and subsequent shock advisory. When attempting a
rescue using the AED, do not operate wireless radiotelephones within 1 meter of
the AED – turn power OFF to the radiotelephone and other like equipment near
the incident.
WARNING: Possible Interference with Implanted Pacemaker.
!
Therapy should not be delayed for patients with implanted pacemakers and a
defibrillation attempt should be made if the patient is unconscious and not
breathing. The AED has pacemaker detection and rejection, however with some
pacemakers the AED may not advise a defibrillation shock. (Cummins, R., ed.,
Advanced Cardiac Life Support; AHA (1994): Ch. 4)
When placing Pads:
- Do not place the pads directly over an implanted device.
- Place the pad at least one inch from any implanted device.
WARNING: Electromagnetic Compatibility.
!
Use of accessories or cables other than those specified, with the exception of
accessories and cables sold by Cardiac Science Corporation as replacement parts
for internal components, may result in increased emissions or decreased
immunity of the AED.
WARNING: Improper Equipment Placement.
!
Position the AED away from other equipment. If it is necessary to use the AED
adjacent to or stacked with other equipment, then observe the AED to verify
normal operations.