5
Data Management
What's in this chapter
◆
Recording Rescue Data
◆
Reviewing Rescue Data
The AED is designed for ease of data management and review. The data
stored in internal memory can be displayed on the PC screen using the
RescueLink software.
Recording Rescue Data
The AED can store up to 60 minutes of ECG monitoring time in the AED's
internal memory. Multiple rescues can be stored in the internal memory,
allowing the rescuer to administer additional rescues without downloading
the data to a PC. Should the internal memory become full, the AED will
purge rescues as needed, beginning with the oldest rescue.
When downloading data, RescueLink will enable the user to select which
rescue to download. See the RescueLink application HELP files for more
information.
Reviewing Rescue Data
To retrieve data from internal memory:
Open the AED lid.
1
Connect the serial cable to the PC and to the AED's serial port under the
2
blue rubber data access cover. The voice prompt will say
"Communications Mode."
Run the RescueLink software program.
3
Select Communications, Get Rescue Data.
4
Select Internal Memory of AED then select OK.
5
Select a rescue by clicking on the date and press OK.
6
Version 112-2025-109 B
Data Management
5-1