The pads have a limited shelf life and should not be used beyond the
expiration date. Keep a fresh, unopened pair of pads plugged into the AED at
all times. Refer to the pad package label for operation temperatures.
An audible and visual alert will indicate after the self-test if the pads are
missing, unplugged, or damaged.

Pad Installation

To install the pads:
Open the lid of the AED.
1
Place the pad package into the lid so that the expiration label is visible
2
through the clear window on the lid. The expiration date of the pads will
then be readable without opening the lid of the AED.
Match the color of the connectors (red to red), then plug the pad
3
connector into the AED case as shown in the drawing. Once the pad
connector is plugged into AED, the PAD indicator should extinguish.
Tuck the excess cable length in the bottom holder as shown in the
4
drawing. With the pad package completely secured to the AED lid, close
the lid.
Make sure the expiration date is visible through the clear window of the
5
lid and check to make sure that the STATUS INDICATOR is GREEN. If
the pads are not installed properly, the STATUS INDICATOR will be
RED. Call Customer Service for assistance.
Caution. Use only Cardiac Science Approved Equipment.
!
Using batteries, pads, cables, or optional equipment other than those approved
by Cardiac Science may cause the AED to function improperly during a rescue.
Version 112-2025-109 B
Getting Started
3-7