Table 13: Parameters (continued)
Parameter
Environmental Conditions Free Fall Drop: IEC 60068-2-32 (1975) AM 2
Shipping and
Transportation Conditions
RHYTHMx ECG Analysis
Performance
7-6
Parameters
Detail
(1990), 1 meter
Bump: IEC 60068-2-29 (1987), 40g and 6000
bumps
Vibration (Random): IEC 60068-2-64 (1993): 10Hz
– 2KHz, 0.005 – 0.0012 g
Vibration (Sine): IEC 60068-2-6 (1995): 10Hz –
60Hz, 0.15 mm and 60Hz – 150Hz, 2g
Enclosure Protection: IEC 60529 (2001), IP24
ISTA Procedure 2A
The AED RHYTHMx ECG Analysis system
analyzes the patient's ECG and advises you when
the AED detects a shockable or non-shockable
rhythm.
This system makes it possible for a person, with
no training in the interpretation of ECG rhythms,
to offer defibrillation therapy to victims of sudden
cardiac arrest.
2
/Hz
Version 112-2025-109 B