Cardiac Science Powerheart AED G3 Plus 9390A Operator's And Service Manual page 25

Hide thumbs Also See for Powerheart AED G3 Plus 9390A:
Table of Contents

Advertisement

Immunity test
Radiated RF
IEC 61000-4-3
70-00914-01 F
1-16
IEC 60601 test level
Compliance level
10 V/m
10 V/m
80 MHz to 2.5 GHz
Electromagnetic emissions standards compliance
Electromagnetic environment—
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the AED, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2
80 MHz to 800 MHz
P
d = 2.3
800 MHz to 2.5 GHz
P
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
b
meters (m)
.
Powerheart® AED G3 Plus 9390A and 9390E