Cardiac Science Powerheart AED G3 Plus 9390A Operator's And Service Manual page 45

Hide thumbs Also See for Powerheart AED G3 Plus 9390A:
Table of Contents

Advertisement

Getting Started
Table 3-2: Preparation (continued)
Voice prompt
"When patient's chest and torso are exposed,
remove square foil package from lid of AED. "
"Tear open foil package across dotted line and
remove pads. "
"Next, separate one of the white pads
completely from blue plastic liner. Begin
peeling from the tabbed corner. "
"Firmly place the pad without the liner on the
patient, exactly as illustrated. This pad can be
placed on either of the two locations shown. "
"Next, peel the blue plastic liner off of the
second white pad. "
"Firmly place the second pad on the opposite
location, exactly as illustrated. "
70-00914-01 F
3-8
Text display
Situation
WHEN CHEST IS BARE
Prompts the rescuer to remove the pads from AED lid.
REMOVE FOIL PACKAGE
TEAR OPEN PACKAGE
Prompts the rescuer to open the pad package and remove pads.
REMOVE PADS
PEEL ONE PAD FROM
Repeats every 3 seconds until the pads are separated. If a pad has
BLUE PLASTIC LINER
been peeled before the prompt starts, this prompt will be skipped.
This prompt will be interrupted when pad is peeled.
PRESS PAD FIRMLY
Prompts the rescuer to place one pad on the patient.
TO CHEST AS SHOWN
PEEL SECOND PAD OFF
Prompts the rescuer to remove the liner from the second pad.
BLUE PLASTIC LINER
PRESS PAD FIRMLY
Repeats until second pad placement is sensed. If the pad is placed
AS SHOWN
before prompt starts then this prompt will be skipped. This prompt
will be interrupted when second pad is placed.