Product Information and Safety
Caution: Lithium Sulfur Dioxide Battery.
Pressurized contents: never recharge, short circuit, puncture, deform,
or expose to temperatures above 65°C (149°F). Remove the battery
when discharged.
Caution: Battery Disposal.
Recycle or dispose of the lithium battery in accordance with all
federal, state and local laws. To avoid fire and explosion hazard, do not
burn or incinerate the battery.
Caution: Use only Cardiac Science Approved Equipment.
Using batteries, pads, cables, or optional equipment other than those
approved by Cardiac Science may cause the AED to function
improperly during a rescue.
Caution: Possible Improper AED Performance.
Using pads that are damaged or expired may result in improper AED
performance.
Caution: Serial Communication Cable.
The AED will not function during a rescue when the serial
communication cable is connected to its serial port. When the serial
communication cable is connected to the AED during a rescue, the
prompt "Remove Cable to Continue Rescue" will be heard until you
remove the serial communication cable.
Caution: Moving the Patient During a Rescue.
During a rescue attempt, excessive jostling or moving of the patient
may cause AEDs to improperly analyze the patient's cardiac rhythm.
Stop all motion or vibration before attempting a rescue.
70-00914-01 F
Powerheart® AED G3 Plus 9390A and 9390E
1-6