Scheduled maintenance
3.
Check the expiration date on the pads.
4.
Check that the battery has adequate charge. If the battery
indicator is red, replace the battery.
5.
Listen for the voice prompts. Additionally, check the display shows
text prompts that correspond to the audio.
6.
Close the lid and observe the change of the STATUS
INDICATOR to RED. After approximately 5 seconds, verify that
the STATUS INDICATOR returns to GREEN.
Annual maintenance
Perform the following tests annually to confirm that the diagnostics are
functioning properly and to verify the integrity of the case.
Check the integrity of the pads and circuitry:
1.
Open the AED lid.
Remove the pads.
2.
3.
Close the lid.
Confirm that the STATUS INDICATOR turns RED.
4.
5.
Open the lid and confirm that the Pad indicator is lit.
Reconnect the pads and close the lid.
6.
7.
Make sure the expiration date is visible through the clear window
of the lid.
8.
Check to make sure that the STATUS INDICATOR is GREEN.
If the pads are not installed properly, the PAD indicator will
illuminate. Contact Cardiac Science Technical Support (see
Contact information
Cardiac Science representative.
9.
Open the lid and confirm that no diagnostic indicators are lit.
10.
Check the expiration date of the pads; if expired, replace them.
11.
Check the pads packaging integrity.
12.
Close the lid
Check the Integrity of the Service Indicator (LED) and Circuitry:
Powerheart® AED G3 Plus 9390A and 9390E
on page 1-2) or outside the U.S., your local
70-00914-01 F
5-5