Cardiac Science Powerheart AED G3 Plus 9390A Operator's And Service Manual page 42

Hide thumbs Also See for Powerheart AED G3 Plus 9390A:
Table of Contents

Advertisement

AED indicators
Text display
The text display has 2 lines of text. The text display provides the
operator with information regarding system initialization, text prompts
and data during a rescue, and diagnostics.
SHOCKS 0
PRESS PAD FIRMLY
System initialization occurs when the lid is first opened. The text
display shows the operator the identifiers for the internal code, voice
prompts and text prompts versions. The text display also shows the
current date and time.
During a rescue, the text display shows the number of shocks delivered
and the elapsed time from the beginning of the rescue (when the lid
was first opened). During CPR, a countdown timer will be displayed.
The text version of the voice prompts will also be displayed.
Note: There is a 3 second delay between the time the AED lid is
opened and the start of the rescue. This 3 second delay is not
included in the elapsed rescue time.
Powerheart® AED G3 Plus 9390A and 9390E
00:20
SHOCKS 0
00:22
AS SHOWN
70-00914-01 F
3-5