Safety alert descriptions
WARNING! Possible Radio Frequency (RF) Susceptibility.
Radio-frequency (RF) interference from devices such as cellular
phones and two-way radios can cause improper AED operation. The
AED should be used at least 6 feet (2 meters) away from RF devices, as
stated in accordance with EN 61000-4-3:2002.
WARNING! Possible Interference with Implanted Pacemaker.
Therapy should not be delayed for patients with implanted
pacemakers and a defibrillation attempt should be made if the
patient is unconscious and not breathing. The AED has pacemaker
detection and rejection, however with some pacemakers the AED
may not advise a defibrillation shock. (Cummins, R., ed., Advanced
Cardiac Life Support; AHA (1994): Ch. 4)
When placing Pads:
WARNING! Electromagnetic Compatibility.
Use of accessories or cables other than those specified, with the
exception of accessories and cables sold by Cardiac Science
Corporation as replacement parts for internal components, may result
in increased emissions or decreased immunity of the AED.
WARNING! Improper Equipment Placement.
Position the AED away from other equipment. If it is necessary to use
the AED adjacent to or stacked with other equipment, then observe
the AED to verify normal operations.
Caution: Restricted Use.
Federal law restricts this device for sale by or on the order of a
physician or practitioner licensed by law of the state in which he/she
practices.
Powerheart® AED G3 Plus 9390A and 9390E
Do not place the pads directly over an implanted device.
Place the pad at least one inch from any implanted
device.
70-00914-01 F