Cardiac Science Powerheart AED G3 Plus 9390A Operator's And Service Manual page 23

Hide thumbs Also See for Powerheart AED G3 Plus 9390A:
Table of Contents

Advertisement

Immunity test
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
61000-4-11
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
U
is the a.c. mains voltage prior to application of the test level.
Note:
T
70-00914-01 F
1-14
IEC 60601 test level
Compliance level
<5% U
Not applicable
T
(>95% dip in U
) for 0.5
T
cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
(30% dip in U
)
T
T
for 25 cycles
<5% U
(>95% dip in U
)
T
T
for 5 sec.
3 A/m
80 A/m
Electromagnetic emissions standards compliance
Electromagnetic environment—
guidance
Power frequency magnetic fields should
be at levels no higher than those
characteristic of a typical location in
typical heavy industrial and power
plants and the control rooms of H.V. sub-
stations.
Powerheart® AED G3 Plus 9390A and 9390E