Technical Data
Table 6-1: Parameters (continued)
Parameter
Cardiac Rhythms Used to
Test the Rhythm
Recognition Detection
System for Powerheart G3
AEDs
70-00914-01 F
6-6
Detail
Shockable Rhythm – VF: Meets IEC 60601-2-4
requirement and AHA recommendation of Sensitivity of
>90%
Automatic External Defibrillators for Public Access
Defibrillation: Recommendations for Specifying and
Reporting Arrhythmia Analysis Algorithm Performance,
Incorporating New Waveforms and Enhancing Safety,
AHA AED Task Force and approved by the AHA Science
Advisory and Coordinating Committee. Circulation,
1997(95), pp 1677-1682
Shockable Rhythm – VT: Meets IEC 60601-2-4
requirement and AHA recommendation of Sensitivity of
>75%
Non-shockable Rhythm – NSR: Meets IEC 60601-2-4
requirement (>95%) and AHA recommendation (>99%)
of Specificity
Non-shockable – Asystole: Meets IEC 60601-2-4
requirement and AHA recommendation of Specificity of
>95%
Non-shockable: Meets IEC 60601-2-4 requirement and
AHA recommendation of Specificity – all other rhythms
of >95%
For detailed information contact Cardiac Science for
white papers:
P/N 112-2013-005 (Pediatric Defibrillation Instructions
for use)
P/N 110-0033-001 (RHYTHMx White Paper)
P/N MKT-11081-01 (STAR Biphasic White Paper)