Parameters
Table 6-1: Parameters (continued)
Parameter
Immunity
Environmental Conditions
Shipping and
Transportation Conditions
RHYTHMx ECG Analysis
Performance
Powerheart® AED G3 Plus 9390A and 9390E
Detail
EM
IEC 61000-4-3, Level X, (20V/m)
IEC 60601-2-4 (20V/m)
Magnetic
IEC 61000-4-8
IEC 60601-2-4
ESD
IEC 61000-4-2
IEC 60601-2-4
6kV contact discharge, 8KV air gap discharge
Free Fall Drop: IEC 60068-2-32, 1 meter
Bump: IEC 60068-2-29, 40g and 6000 bumps
Vibration (Random): IEC 60068-2-64: 10Hz – 2kHz, 0.005 –
2
0.0012 g
/Hz
Vibration (Sine): IEC 60068-2-6: 10Hz – 60Hz, 0.15 mm
and 60Hz – 150Hz, 2g
Enclosure Protection: IEC 60529, IP24
Vibration (random): RTCA DO-160D Section 8, category S,
curve B
Temperature variation: RTCA DO-160D Section 5,
category C
Temperature/altitude decompression/overpressure:
RTCA DO-160D section 4, category A4, operating 0ºC to
50ºC, ground survival 0ºC to 50ºC
ISTA Procedure 2A
The AED RHYTHMx ECG Analysis system analyzes the
patient's ECG and advises you when the AED detects a
shockable or non-shockable rhythm.
This system makes it possible for a person, with no
training in the interpretation of ECG rhythms, to offer
defibrillation therapy to victims of sudden cardiac arrest.
With a new battery, after the AED has delivered 15 300VE
shocks, the maximum time from beginning rhythm
analysis until the AED is ready to shock is 17 seconds.
70-00914-01 F
6-5