Technical Data
Table 6-2: Ultra-low Variable Energy (150 VE) Powerheart G3 Waveform
Phase 1
Patient's
Impedance
Voltage
(Ohms)
* (Volts)
25
1393
50
1420
75
1430
100
1434
125
1437
150
1439
175
1441
Table 6-3: Low Variable Energy (200 VE) Powerheart G3 Waveform
Phase 1
Patient's
Impedance
Voltage
(Ohms)
* (Volts)
25
1609
50
1640
75
1651
100
1656
125
1660
150
1662
175
1663
70-00914-01 F
6-8
Phase 2
Duration*
Voltage
(MS)
* (Volts)
3.3
743
4.5
909
5.8
973
7.0
1007
8.3
1027
9.5
1040
10.8
1049
Phase 2
Duration*
Voltage
(MS)
* (Volts)
3.3
858
4.5
1050
5.8
1124
7.0
1163
8.3
1186
9.5
1201
10.8
1212
Powerheart® AED G3 Plus 9390A and 9390E
Duration*
Energy**
(MS)
(Joules)
3.2
145-196
3.2
128-173
3.2
116-156
3.2
108-146
3.2
102-138
3.2
98-132
3.2
95-128
Duration*
Energy**
(MS)
(Joules)
3.2
193-260
3.2
170-230
3.2
155-209
3.2
144-194
3.2
136-184
3.2
130-176
3.2
126-170