STAR biphasic waveform
Table 6-4: High Variable Energy Powerheart G3 Waveform (all values are
typical)
Phase 1
Patient's
Impedance
Voltage
(Ohms)
* (Volts)
25
1869
50
1906
75
1918
100
1925
125
1928
150
1931
175
1933
* All values are typical.
**Allowable energy range.
Powerheart® AED G3 Plus 9390A and 9390E
Phase 2
Duration*
Voltage
(MS)
* (Volts)
3.3
997
4.5
1220
5.8
1306
7.0
1351
8.3
1378
9.5
1396
10.8
1408
Duration*
Energy**
(MS)
(Joules)
3.2
260-351
3.2
230-311
3.2
209-283
3.2
195-263
3.2
184-248
3.2
176-238
3.2
170-230