This section presents information about the AED diagnostics self-tests,
maintenance, and service indications.
Self-tests
The AED has a comprehensive self-test system that automatically tests
the electronics, battery, pads, and high voltage circuitry. Self-tests are
also activated every time you open and close the AED lid.
When performing the self-tests, the AED completes the following steps
automatically:
1.
Turns itself on, and the Status Indicator changes to red.
2.
Performs the self-test.
3.
If successful, the Status Indicator reverts to green.
4.
Turns itself off if the lid is closed.
There are three types of automatic self-tests:
◆
The daily self-test checks the battery, pads, and the electronic
components.
Powerheart® AED G3 Plus 9390A and 9390E
Troubleshooting and
Maintenance
Self-tests
Indicator troubleshooting table
Scheduled maintenance
Authorized repair service
Frequently Asked Questions
5-1
5-3
5-4
5-6
5-7
70-00914-01 F
5-1