Cardiac Science Powerheart AED G3 Plus 9390A Operator's And Service Manual page 51

Hide thumbs Also See for Powerheart AED G3 Plus 9390A:
Table of Contents

Advertisement

Getting Started
Table 3-8: Other prompts (continued)
Voice prompt
Text display
"Rhythm changed.
RHYTHM CHANGED
Shock cancelled. "
SHOCK CANCELLED
"Remove cable to
REMOVE CABLE
continue rescue. "
TO CONTINUE RESCUE
"Communications
COMMUNICATIONS MODE When the lid is open and the serial communication cable is plugged into the AED.
mode"
"Service required"
SERVICE REQUIRED
70-00914-01 F
3-14
Situation
When the device is prepared to shock then detects a change in rhythm and therefore
cancels the shock.
When a serial communication cable is connected to the AED during a rescue, the phrase
repeats until the cable is disconnected.
Occurs after the self-tests determine that the AED is not functioning properly. The
prompt "SERVICE REQUIRED" will be heard when the lid is opened. The red SERVICE
indicator will illuminate. After closing the lid, an alarm beep will be heard until the
battery is removed or becomes completely depleted.
Powerheart® AED G3 Plus 9390A and 9390E

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Powerheart aed g3 plus 9390e

Table of Contents