CM4 Four-Point Continuous Monitor
9.1 Introduction
The Maintenance Section includes:
•
Chemcassette
Maintenance covers the Chemcassette
®
Chemcassette
carrier, optics verification, and flowrate check
®
(Section 9.2)
•
Filter maintenance covers the replacement of CM4 filters
(Section
9.3)
•
Fuse replacement
(Section
•
Cleaning Chemcassette
•
Installing Microtube assemblies
NOTE:
Prior to performing any maintenance procedure, leave Monitor
Mode. (See
Section
4.5.1.)
CAUTION
For system integrity, always perform a leak test after any maintenance
procedure. See
Section B.2
9.2 Chemcassette
Maintenance
®
This section describes maintenance and handling procedures for the
Chemcassette
and the Chemcassette
®
Perform the following functions when replacing a Chemcassette
tape:
•
Remove carrier
•
Verify optics
•
Load new Chemcassette
•
Install Chemcassette
®
•
Check flowrates
CM4 Technical Manual
9.4)
optics
(Section
9.5)
®
(Section
9.6)
for leak check procedure.
carrier.
®
in carrier
®
9.2.1 Remove Carrier
To remove the Chemcassette
1. Press PROGRAM to exit Monitor Mode.
,
®
2. From the 4–SERVICE MENU, Choose the subfunction 5–LOAD
Chemcassette
.
®
5 – LOAD Chemcassette
If Chemcassette
counter is disabled, proceed to step 6. If
®
Chemcassette
counter is enabled, the screen displays:
®
3. Press either:
•
1 to re-initialize the counter so that the display shows a fault
when less than 24 hours of Chemcassette
•
0 to continue without resetting counter
If the internal filter timer is active (see
displays:
INT FILTER RESET 1=Y
4. Press either:
•
1 to reset the timer after servicing the filters (see
9.3)
®
•
0 to continue without resetting timer
If the external filter timer is active (see
displays:
EXT FILTER RESET 1 = Y
9-2
carrier from the CM4:
®
4 – SERVICE MODE
RESET COUNTER 1=Y
remains
®
or
Section
7.8.2), the screen
or
Section
7.8.3) , the screen
Section