CM4 Four-Point Continuous Monitor
2 – REVIEW MODE
2 – FAULT HISTORY
3. Press 1 to display the fault history.
F12 END OF CASSETTE
08/26/05 10:30
The display shows the last eight faults stored in fault history.
4. Press either:
CURSOR ARROW LEFT/RIGHT to scroll forward/backward
to the next/previous event stored
ENTER to repeat the event just displayed
Fault history is not cleared using the RESET key. Press RESET to
return to the Review Menu.
6.4 Function 2.3 – Current TWA
Use the Current TWA (Time Weighted Average) function to display or
print the current TWA for each point monitored since the start of the
current 8-hour TWA. The TWA is the concentration of gas measured
and averaged over the entire period.
6.4.1 Print TWA
See
Figure 6-4
for an example of a typical TWA printout.
Purpose
The printout of the TWA shows:
Instrument serial number
Current date and time
Customer name and location
Start and end time
CM4 Technical Manual
or
Elapsed time
Concentration levels for each point
Instructions
1. Press PROGRAM to exit Monitor Mode.
2. From the 2–RE VIE W MENU, choose the subfunc tion
3–CURRENT TWA.
2 – REVIEW MODE
3 – CURRENT TWA
The display reads:
1 = DISPLAY 2 = PRINT
3. Press 2 to print the TWA history.
The CM4 calculates and prints TWA information based upon the
elapsed time (maximum of 8 hours) from the last entry into Monitor
Mode, power-up, or completed TWA printout until the last sample prior
to leaving Monitor Mode.
08/26/05 13:51 CM4-851-0000 TWA
ABC MANUFACTURING CO., CHICAGO, IL
START TIME:
END TIME:
ELAPSED TIME:
POINT 1 GASCAB
POINT 2 VALVEBOX 8 PPB
POINT 3 OPZONE
POINT 4 EXHDUCT 0 PPB
Figure 6-4: TWA Printout
6-5
08/26/05 03:56
08/26/05 11:26
7 HOURS 30 MINUTES
16 PPB
AsH3
AsH3
0 PPB
AsH3
AsH3