Honeywell CM4 Technical Handbook page 135

Four-point continuous monitor
Hide thumbs Also See for CM4:
Table of Contents

Advertisement

CM4 Four-Point Continuous Monitor
Gas Name
Hydrogen Bromide XP4
(HBr)
Hydrogen Chloride
(HCl)
Hydrogen Chloride XP
(HCl)
Hydrogen Chloride XP4
(HCl)
Hydrogen Chloride XP4
(HCl)
Low Level
Hydrogen Chloride
(HCl)
Low Level
Hydrogen Chloride XP
(HCl)
Low Level
* = Minimum Sample Time TLV = Threshold Limit Value LAL = Lowest Alarm Level LDL = Lower Detectable Limit
CM4 Technical Manual
Default
TLV
LAL
LDL
Alarm
Level 1
2 ppm
0.3 ppm 0.2 ppm
2 ppm
2 ppm
0.5 ppm 0.5 ppm
2 ppm
2 ppm
0.2 ppm 0.2 ppm
2 ppm
2 ppm
0.2 ppm 0.2 ppm
2 ppm
2 ppm
30 ppb
20 ppb
200 ppb 400 ppb 20-2000 ppb
2 ppm
100 ppb
80 ppb
500 ppb
2 ppm
30 ppb
20 ppb
200 ppb 400 ppb 20-2000 ppb
Table C-1: Chemcassette Detectable Gases (cont'd)
Default
Alarm
Range
Alarm Setting
Level 2
0.3-1.9 ppm
4 ppm
0.2-20 ppm
2-20 ppm
0.5-9.9 ppm
4 ppm
0.5-15 ppm
10-15 ppm
0.2-0.9 ppm
1-3.9 ppm
4 ppm
0.2-20 ppm
4-20 ppm
0.2-0.9 ppm
1-3.9 ppm
4 ppm
0.2-20 ppm
4-20 ppm
30-99 ppb
100-199 ppb
200-399 ppb
400-2000 ppb
100-499 ppb
500-999 ppb
1000
80-8000 ppb
1000-2999 ppb
ppb
3000-8000 ppb
30-99 ppb
100-199 ppb
200-399 ppb
400-2000 ppb
C-7
Time to 1 TLV
Initial
Alarm @ 2 TLV
Analysis
Chemcassette
Concentration
Period
Part Number
10 Foot Sample
(seconds)
Line
45
30
<20 sec
1258-9310
5*
30
20
<20 sec
874337
0*
60
30
<15 sec
1874-9310
20
0*
60
30
<15 sec
1258-9310
20
0*
300
<50 sec
150
(Alarm @ 200
60
1258-9310
ppb with 400
30
ppb HCl gas)
0*
300
120
60
<20 sec
874337
30
5*
300
<50 sec
150
(Alarm @ 200
60
1874-9310
ppb with 400
30
ppb HCl gas)
0*

Hide quick links: