Honeywell CM4 Technical Handbook page 134

Four-point continuous monitor
Hide thumbs Also See for CM4:
Table of Contents

Advertisement

CM4 Four-Point Continuous Monitor
Gas Name
Germane XP4
(GeH
)
4
Hydrazine
(N
H
)
2
4
Hydrogen Bromide
(HBr)
Hydrogen Bromide XP
(HBr)
Hydrogen Bromide XP4
(HBr)
Low Level
Hydrogen Bromide
(HBr)
Low Level
Hydrogen Bromide XP
(HBr)
Low Level
* = Minimum Sample Time TLV = Threshold Limit Value LAL = Lowest Alarm Level LDL = Lower Detectable Limit
CM4 Technical Manual
Default
TLV
LAL
LDL
Alarm
Level 1
200 ppb 100 ppb 100 ppb 200 ppb 400 ppb
10 ppb
10 ppb
10 ppb
100 ppb 200 ppb 10-1000 ppb
2 ppm
0.5 ppm 0.3 ppm
2 ppm
2 ppm
0.3 ppm 0.2 ppm
2 ppm
2 ppm
30 ppb
20 ppb
200 ppb 400 ppb 20-2000 ppb
2 ppm
100 ppb
50 ppb
500 ppb
2 ppm
30 ppb
20 ppb
200 ppb 400 ppb 20-2000 ppb
Table C-1: Chemcassette Detectable Gases (cont'd)
Default
Alarm
Range
Alarm Setting
Level 2
100-149 ppb
150-199 ppb
100-2000
200-2000 ppb
ppb
10-14 ppb
15-19 ppb
20-1000 ppb
0.5-2.9 ppm
3-5.9 ppm
4 ppm
0.3-30 ppm
6-8.9 ppm
9-30 ppm
0.3-1.9 ppm
0.2-20 ppm
2-20 ppm
4 ppm
30-99 ppb
100-399 ppb
400-2000 ppb
100-249 ppb
250-499 ppb
1000
50-2000 ppb
500-749 ppb
ppb
750-2000 ppb
30-99 ppb
100-399 ppb
400-2000 ppb
C-6
Time to 1 TLV
Initial
Alarm @ 2 TLV
Analysis
Chemcassette
Concentration
Period
Part Number
10 Foot Sample
(seconds)
Line
480
360
<240 sec
1258-9300
240
5*
600
360
874347
<360 sec
240
5*
45
30
15
<20 sec
874337
10
2*
45
30
1874-9310
<20 sec
5*
180
<60 sec
120
(Alarm @ 500
1258-9310
60
ppb with 1000
5*
ppb HBr gas)
240
<60 sec
120
(Alarm @ 500
60
874337
ppb with 1000
30
ppb HBr gas)
5*
180
<60 sec
120
(Alarm @ 500
1874-9310
60
ppb with 1000
5*
ppb HBr gas)