Defining The Point Id - Honeywell CM4 Technical Handbook

Four-point continuous monitor
Hide thumbs Also See for CM4:
Table of Contents

Advertisement

CM4 Four-Point Continuous Monitor
9. Press ENTER to accept the value and advance to the display
prompt used to set the current loop Full Scale values:
20 mA = 500 PPB
Setting the current loop Full Scale values applies to the 4-20 mA
current loop. To select the 2 mA fault indication, see Section 7.4.5.
The display reads:
20 mA = 500 PPB
The cursor flashes over the Full Scale (FS) value.
10. Press CURSOR ARROW LEFT/RIGHT and numeric keys to
enter the desired full scale value.
11. Press ENTER to accept the value and advance to the following
display:
1=GAS/LEVELS 2=PT ID
or
Press RESET to select another point. The display reads:
SELECT POINT
The flashing status LED stops. Press RESET again to display the
print prompt:
PRINT? 1=YES 0=NO
12. Repeat Steps 4 through 11 to set the gas type and alarm for
the remaining three points.
13. After programming the chosen points, press ENTER to accept
the values. The display reads:
1=GAS/LEVELS 2=PT ID
Press 2 to define the point identification (see Section 7.3.2).
CM4 Technical Manual

7.3.2 Defining the Point ID

This feature permits changing the default point ID label to a more
informative string. The name will be displayed on the front panel display
during monitor mode.
Instructions
1. Press PROGRAM to exit Monitor Mode.
2. From the 3–SET-UP MENU, choose the subfunction 2–POINT
PARAMETERS.
2 – POINT PARAMETERS
The display reads:
3. Use the numeric keys to specify the point you wish to configure,
or press RESET to exit this function.
If printer output is enabled, the display reads:
PRINT? 1=YES 0=NO
The printout helps to determine which point parameters to change.
If, for example, you selected point 1 (pressed 1) in Step 3, the printer
prints the parameters and the display briefly reads:
POINT 1 SELECTED
The corresponding point status LED blinks.
The display changes to:
1=GAS/LEVELS 2=PT ID
4. Press 2 to define the point identification.
The display briefly reads:
7-5
3 – SET-UP MODE
SELECT POINT
PROGRAM RESET