Introduction - Honeywell CM4 Technical Handbook

Four-point continuous monitor
Hide thumbs Also See for CM4:
Table of Contents

Advertisement

CM4 Four-Point Continuous Monitor

8.1 Introduction

The Service Menu provides access to a variety of diagnostic tests
and routine service procedures. The Service Menu contains eight
functions.
4
SERVICE MENU
4.1
RELAY/LED TEST
4.2
ALARM TEST
4.3
DIAGNOSTICS
4.4
RESET SYSTEM
4.5
LOAD Chemcassette
4.6
SET FLOW
4.7
MDA SERVICE
4.8
PUMP LIMITS
Figure 8-1: Service Mode Menu
The following is an overview of each of the eight functions. The following
sections describe each of these functions in detail:
4.1 RELAY/LED TEST checks all LED indicators and alarm
relays (see Section 8.2)
4.2 ALARM TEST checks alarm LEDs and alarm relays (see
Section
8.3)
4.3 DIAGNOSTICS prints out operational information about
the CM4 (see
Section
4.4 RESE T SYSTEM deletes all operator-programmed
settings and replaces them with the factory default settings
(see
Section
8.5)
4.5 LOAD Chemcassette
reset the tape counter after loading a new Chemcassette
(see
Section
8.6)
CM4 Technical Manual
8.4)
/FILTER CHANGE, if selected, will
®
4.6 SET FLOW balances the sample flowrates at each point
from all CM4 sample lines through the system (see
8.7)
4.7 MDA SERVICE for use by MDA service technicians only
(see
Section
8.8)
4.8 PUMP LIMITS (see
8.2 Function 4.1 – Relay/LED Test
Purpose
Use the Relay/LED Test to test all LED indicators and all relays except
the Watch Dog relay.
NOTE:
The relay test simulates an actual alarm condition and the CM4
activates the appropriate connected relays (including point alarms
and general alarms). Notify appropriate personnel that you plan to
conduct a test.
Instructions
1. Press PROGRAM to exit Monitor Mode.
2. From the 4–SER VICE MENU, choose the subfunc tion
1–RELAY/LED TEST.
1 – RELAY/LED TEST
Press 1 again to turn on all fault and alarm relays and all LEDs
Press ENTER to choose either LEDs or relays. The display reads:
1=LEDs ONLY 2=RELAYS
3. Press 1 to light all LEDs.
®
Press 2 to activate all output relays. The display reads:
8-2
Section
8.9)
4 – SERVICE MODE
or
CONFIRM TEST
Section

Hide quick links:

Advertisement

Table of Contents