Display of the parameter concerning the sensitivity used on the clamp
The range displayed on the screen of the Model 8220 must be consistent with the
sensitivity used on the SL261 probe 10A for the 10mV/A sensitivity (Figure 4-6)
and 100A for the 100mV/A sensitivity (Figure 4-7).
The value is selected using the White and Yellow (
values are 10 and 100 (10mV/A and 100mV/A).
Parameter
Page 5:
The
symbol means event per rotation.
This parameter is designed to configure the rotation speed mode (
The
parameter defines the number of events per rotation for the measure-
ment of rotation speed for a machine in operation.
Example: If a tachometer signal provides two pulses per revolutions, this parameter
would be set to 2.
The value is set using the White and Yellow (
maximum values are 1 and 99.
Page 6:
The
symbol means threshold.
This parameter is designed to configure the rotation speed mode (
Power Quality Analyzer Model 8220
Parameter
Figure 4-7
) buttons. The minimum and
) buttons. The preset
).
).
33