AEMC 8220 User Manual page 24

Single-phase power quality analyzer
Table of Contents

Advertisement

Function
Displacement factor
(DPF)
Level of harmonics
order ∈ [1; 50]
(V RMS > 50V)
Without AmpFlex
or MiniFlex
®
(I RMS > 3 × I nom ÷ 100)
AmpFlex
or MiniFlex
®
(I RMS > I nom ÷ 10)
Global level of harmonics
(THD-F) order ≤ 50
Distortion factor (THD-R)
order ≤ 50
K factor
Rotation Speed
Temperature
Resistance
(1) 1.2 x 1000 x √2 - 1700A
(2) 1.2 x I
x √2 = 1.7 x Inom
nom
(3) 6500 x √2 - 9190A
(4) 600 x √2 = 850V
(5) For the phase-to-earth voltage measurement (phase-neutral). For the phase-to-phase voltage measurement
(phase-phase), in balanced three-phase mode, it is possible to reach 660V
having a phase-neutral voltage of 380V
(6) An additional influence of 3.5°C must be added in an environment disturbed by radiation fields. In other words,
the error in the field of reference in an environment disturbed by radiation fields is ± (1 %+4.5°C).
An additional influence of 6.2°F must be added in an environment disturbed by radiation fields. In other words,
the error in the field of reference in an environment disturbed by radiation fields is ± (1 %+8°F).
NOTE: The uncertainties given for power and energy measurements are maximum for Cos φ = 1 or Sin φ = 1 and
are typical for the other phase shifts.
22
Measurement range
Minimum
-1
*
0%
®
*
®
0%
0%
1
6 RPM
-200.0 °C
-328.0 °F
0.0Ω
).
RMS
Display resolution
Maximum
1
0.001
999.9%
0.1%
999.9%
0.1%
999.9%
0.1%
99.99
0.01
0.1 RPM
V < 1 kRPM
1 RPM
1 kRPM ≤ V < 10 kRPM
120 kRPM
10 RPM
10 kRPM ≤ V
< 100 kRPM
100 RPM
V ≥ 100 kRPM
850.0 °C
0.1 °C
0.1 °F
T < 1000°F
1562 °F
1 °F
T ³ 1000°F
1 Ω
2000Ω
R < 1000Ω
Power Quality Analyzer Model 8220
Accuracy
±(1°) on φ
±(5cts) on DPF
±(1%+5cts)
±(1%+5cts)
±(1%+10cts)
±(5%)
± (0.5 %)
± (1 % + 1 °C)
± (1.8 % + 2 °F)
± (1.5 %+2 cts)
(balanced three-phase network
RMS
(6)
(6)

Advertisement

Table of Contents
loading

Table of Contents