The Configure dialog box consists of the following features:
•
Reactive Values Calculation: With or without harmonics. Applies to VAR
calculation.
•
Nominal Frequency: 50 or 60Hz (displays last detected frequency).
•
Current Sensors: MN93, MN193, SR193, MR193, AmpFlex
5A Adapter or SL261. Probes displayed are automatically detected by the
instrument. (See § 4.3 for configuration of the MN193 and Adapter box).
•
Connection Type: Single-Phase or 3-Phase balanced.
•
Voltage Transformer Ratio: Sets the scale for voltage measurement in
cases where measurements are on the secondary side of a transformer and
the primary value needs to be displayed. Data is saved on the PC, but not
written to the instrument. Corrected values will only be visible in DataView
•
Instrument Clock: Programs the computer's time and date into the
configuration of the Model 8220.
NOTE: For detailed instructions and descriptions for any feature in a dialog box,
right-click on the feature you want information about.