The value is set using the White and Yellow (
are 0, 1, 2, 5 and 10%.
NOTE: The effective half-cycle stopping value is calculated using the following
formula: [effective half-cycle stopping value] = [effective half-cycle starting
value] × (100 – [hysteresis]) ÷ 100.
Parameter
Page 3:
The
symbol means primary.
This parameter configures the primary current of the transformation ratio (
NOTE: This screen is only displayed if the Model 8220 is connected to a 5A
adapter or to the MN193 5A clamp. For all of the other sensors, which do not
require any adjustment, this screen is not displayed:
The
parameter defines the value of the primary effective current for the MN193
clamp (5A) or the 5A adapter.
The value is set using the White and Yellow (
maximum terminals are 0 and 2999A.
Power Quality Analyzer Model 8220
Figure 4-4
Display of the effective primary current.
) buttons. The preset values
) buttons. The minimum and
).
31