64 Disassembly and repair
6. Remove the AC power cord.
7. Remove the comms connection cover.
Loosen the two captive screws, using a number 2 Phillips screwdriver, until the
cover is free from the housing.
8. Remove the wall anchor screws using a number 2 Phillips screwdriver.
9. Hold the bottom and top of the system and push it up to remove the system from
the wall mounting bracket.
10. Set the system upside down on a large, flat surface. Make sure that the display is
protected from pressure or getting scratched.
11. Remove the battery.

Remove the handle module assembly

Remove the system from the wall and remove the battery as described.
1. Place the device on a large, flat surface with the display facing down.
Ensure that the display is protected from pressure or getting scratched.
2. Cut and remove the cable tie that holds the USB ferrite in place.
3. Remove the ferrite from the USB cable by pulling apart the two clips and lifting up.
®