Service manual
3. Disconnect the power button cable from the main board at J6.
4. Disconnect and remove the Ethernet cable.
5. Remove the four large Phillips screws that mount the LCD to the housing.
6. Carefully lift the LCD and the attached main board from the housing.
7. Remove the module well as described.
8. Remove the two module tracks by removing the four torx screws.
CAUTION Support the LCD as you lift the assembly from
the housing. The LCD is not secured to the assembly.
Disassembly and repair 85