62 Disassembly and repair
3. For systems with the Braun thermometer option, disconnect the USB cable from the
rear of the Braun housing.
4. Locate and remove the module retaining plate by removing the four #2 Phillips
screws.
5. Slide any blank face plates or modules to be removed toward the open end.
Note
Note
Remove the Braun dock from the housing
Hold the Braun housing with your right hand and the dock with your left hand as shown.
With your right-hand fingers inside the housing, gently press the housing toward you and
away from the dock. At the same time, gently push the dock away from you until the
dock unsnaps from the housing. When the dock is completely loose, remove it from the
housing.
When removing any connected modules, disconnect the
USB cables before you remove the module.
If you are replacing a Braun 4000 thermometer with a Braun
6000 thermometer, you must also replace the dock.
®