54 Troubleshooting
Message
Information
Device is operating in battery
mode.
Configuration Manager messages
Message
Alarm
Unable to load configuration; using
factory defaults.
Functional error. Call for service.
Information
No connection for send.
Patient data management messages
Message
Alarm
Maximum number of patient
records saved. Oldest record
overwritten.
Information
No data saved.
Patient ID required to save data.
Clinician ID required to save data. The configuration requires a clinician ID to save
Possible cause
The AC power cord has been disconnected
Possible cause
A configuration load error occurred
A critical configuration load error occurred
The monitor is not configured to the network
Possible cause
The maximum number of patient records has
been exceeded
No patient data is available
The configuration requires a patient ID to save
data
data
®
Welch Allyn Connex
Integrated Wall System
Suggested action
Touch the alarm icon to dismiss
or plug the monitor to AC power.
Suggested action
Restore factory defaults. If the
error persists, replace the main
board.
Replace the main board.
Change advanced setting to
Automatically
disable
send on manual save
control.
Configure the monitor to the
network.
Suggested action
Go to the Review tab and delete
old records to prevent the alarm
from appearing when new
records are saved.
Take or enter vital signs before
saving.
Require patient
Disable
ID to save readings
Patient
the
tab, available from
Data management
the
tab in Advanced settings.
Require clinician
Disable
ID to save readings
Clinician
the
Data
from the
on
on
tab, available