Welch Allyn Connex Service Manual page 95

Integrated wall system
Hide thumbs Also See for Connex:
Table of Contents

Advertisement

Service manual
Full functional verification check and calibration tools
The list of tools below is what is required to perform a full device functionality check and
calibration. The tools are used in conjunction with the Welch Allyn service tool, Gold
licensed edition, to perform a device calibration.
Material no.
01802-110
407672
411690
N/A
DOC-10
06138-000
N/A
N/A
4500-925
106270
660-0138-00
407022
N/A
6000-30
N/A
N/A
web download
N/A
Description
Tester, calibration, 9600 Plus
BP test volume repair fixture 113670
PRO 6000 calibration fixture*
*Use with older models have a dust cover screw that rises
above the O-ring washer. Newer models have a dust cover
screw that is recessed into the washer.
Masimo Rainbow SET tester (Masimo part number 2368)
Cable, SpO2 extension, Nellcor
Cal-key, assembly, M690/692
Pressure meter (must include at least one decimal point and
be accurate to within ±0.5 mmHg)
Nellcor portable oximetry tester (SRC-MAX)
USB 2.0/5-pin Mini-B cable, gold, 6 ft.
Service test box
Cable, patch 5 feet RJ45, T568B
Assy, MMF Nurse call cable - service
USB cable Type A to Type B (3 ft.)
Single tube blood pressure hose, 5 ft.
Netgear wireless router, a/b/g or equivalent
PC running Windows XP SP3, Windows 7, or Windows 8
Welch Allyn Service Tool (version 1.8 or later)
Click this link
Blood pressure Y-tube
Functional verification and calibration 91
Qty
Component
3
Braun ear and
SureTemp
Thermometry Probe
1
NIBP
3
Braun ear
1
Masimo SPO2
1
Nellcor SPO2
1
SureTemp
Thermometry Module
1
NIBP
1
Nellcor SPO2
1
NIBP
1
Nurse Call, Voltage &
Current,USB ports
1
Ethernet Test
1
Nurse Call (Service
Test Box Required)
1
USB Ports (Service
Test Box Required)
1
NIBP
1
Wired & Wireless
Test
1
All
1
NIBP, Software
updates
1
NIBP

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents