Calibration

Chapter IV

Calibration Setup

The Thermo Scientific Orion 2120XP calcium hardness analyzer provides
simple and fast calibration cycles using double known addition that quickly
have your system back online with security and confidence. The 2120XP
analyzer monthly calibration is quick and accurate using a simple pipet to
inject two known standards into the sample matrix.
Prior to performing a calibration, refer to Chapter III, Use of the Setup
Mode for instructions on how to properly configure the 2120XP analyzer.
The value of the calibration standards, volume of the calibration standards
and the volume of the flow cell must be entered in the setup mode.
Adjust the flow cell volume according to the precise factory calibration
volume that is printed on the side of each flow cell.
Thermo Scientific Orion 2120XP Calcium Hardness Analyzer User Guide
IV-1