Analyzer Operation
Setup Mode Overview
General Setup Mode
Menu Options
Thermo Scientific Orion 2120XP Calcium Hardness Analyzer User Guide
The setup mode features programmable menu options. The order of the
menu options is dependent on the direction the operator scrolls by pressing
or
The following menu options are displayed in the main setup mode of one
and two channel analyzers.
PASS
Set either of two password options:
•
System password – Setup settings protected, accessed by authorized
operators only
•
Calibration password – Calibration menu data is protected, accessed by
authorized operators only
•
Default password is 0000 – Disables both passwords
•
Forgot your password? Contact Technical Support at 1-800-225-1480
DATE
Set the date in US or Europe format:
•
Enter month, day and year
•
Default date – System will continue to keep date and time due to
battery back up, operator must set in accordance to local time zone
•
If the battery is removed, the system will show 01/01/2000
TIME
Set the time:
•
Enter hour and minutes in 24 hour format
•
Default time – System will continue to keep date and time due to
battery back up, operator must set in accordance to local time zone
•
If the battery is removed the system will show 00:01
LOG
Set the data logging interval for measure log (calibration and error logs are
accessed in the test mode):
•
Set the log interval as hour:minute
•
Default log interval is 00:00 – logging disabled
•
Minimum log interval is 1 minute, maximum log interval is 99 hours
and 59 minutes