Calibration
Offline Calibration
Procedure
Step
Operator Action
1
Open the bypass/needle valve to redirect
the sample flow.
2
Press
to initiate the calibration.
3
If the calibration password is active, enter
the calibration password by pressing
/
to set the first digit,
to move to the next digit,
/
to set the values of the
remaining digits and
through the remaining digits.
Press
to accept the password.
4
5
Press
/
channel that will be calibrated and press
to begin the calibration.
6
7
Press
/
and press
to accept the setting.
8
Wait 1 to 10 minutes.
9
Take a QC sample from the 2120XP
analyzer bypass for laboratory analysis.
Press
.
Thermo Scientific Orion 2120XP Calcium Hardness Analyzer User Guide
The offline calibration feature is simply a one point calibration without
a slope change. The term "offline calibration" refers only to the fact that
a sample from 2120XP analyzer bypass is taken "offline" for laboratory
analysis; in fact, no downtime is experienced and the analyzer remains
online throughout the procedure.
Scrolling Marquee
Actual temperature reading
ENTER PASSWORD
to move
CH1 or CH2
(flashing)
CH1 or CH2
to select the
(flashing)
DKA or OFFL
(flashing)
OFFL
to select OFFL
(flashing)
MEASURING SAMPLE
WAIT 1 MINUTE
AWAITING DATA STABILITY
TAKE SAMPLE FROM BYPASS DRAIN
PRESS ENTER TO USE METER
THEN PRESS CAL TO ENTER LAB
RESULTS
Main Display
Notes
Actual concentration
reading
HOLD and CAL appear
as mode.
PASS
# # # #
CAL
Displayed for two
channel analyzer
CHnL
CAL
Action required for
two channel analyzer
CHnL
CAL
tYPE
CAL
tYPE
Actual readings
System continuously
flashing
evaluates the
data until a stable
condition is met.
LAb
IV-15