Interface Property; Process Property - Endress+Hauser Levelflex FMP55 Operating Instructions Manual

Guided wave radar
Hide thumbs Also See for Levelflex FMP55:
Table of Contents

Advertisement

Operating menu

Process property

Navigation
Description
Selection
Factory setting
Additional information

Interface property

Navigation
Prerequisite
Description
Selection
Factory setting
124
"Interface" submenu
Navigation
 Setup → Advanced setup → Interface
 Setup → Advanced setup → Interface → Process property
Specify typical rate of change for the interface position.
• Fast > 1 m (40 in)/min
• Standard < 1 m (40in) /min
• Medium < 10 cm (4in) /min
• Slow < 1 cm (0.4in) /min
• No filter / test
Standard < 1 m (40in) /min
The device adjusts the signal evaluation filters and the damping of the output signal to the
typical rate of level change defined in this parameter:
Process property
Fast > 1 m (40 in)/min
Standard < 1 m (40in) /min
Medium < 10 cm (4in) /min
Slow < 1 cm (0.4in) /min
No filter / test
 Setup → Advanced setup → Interface → Interface prop.
Operating mode (→  111) = Interface with capacitance
Select interface property.
The interface property determines how the Guided Radar and the Capacitance
Measurement interact.
• Special: automatic DC
• Build up
• Standard
• Emulsion layer
Standard
Levelflex FMP55 HART
Step response time / s
5
15
40
74
2.2
Endress+Hauser

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Levelflex FMP55 and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Levelflex fmp54Levelflex fmp52

Table of Contents