Measuring cell
Influence of the sys-
tem pressure to the
zero point
Influence of the sys-
tem pressure to the
span
Measuring cell
Influence of the system
pressure to the zero point
Influence of the system
pressure to the span
Tantalum diaphragm
Measuring cell
Influence of the sys-
tem pressure to the
zero point
Influence of the sys-
tem pressure to the
span
Measuring cell
Influence of the system
pressure to the zero point
Influence of the system
pressure to the span
Total accuracy
Total Performance - Basic version
The specification "Total Performance" comprises non-linearity incl. hysteresis and non-repeat-
ability, thermal change of the zero point and static pressure influence (p
Total Performance
VEGADIF 65 • Foundation Fieldbus
10 mbar
30 mbar
±0.15 % of URL/
±0.35 % of URL/
7 bar
70 bar
±0.035 % of URL/
±0.14 % of URL/
7 bar
70 bar
3 bar
±0.075 % of URL/7 bar
±0.14 % of URL/7 bar
10 mbar
30 mbar
±0.28 % of URL/
±0.70 % of URL/
7 bar
70 bar
±0.28 % of URL/
±0.70 % of URL/
7 bar
70 bar
3 bar
±0.14 % of URL/7 bar
±0.14 % of URL/7 bar
100 mbar
±0.15 % of URL/
70 bar
±0.14 % of URL/
70 bar
16 bar
±0.075 % of URL/70 bar
±0.14 % of URL/70 bar
100 mbar
±0.42 % of URL/
70 bar
±0.42 % of URL/
70 bar
16 bar
±0.14 % of URL/70 bar
±0.14 % of URL/70 bar
11 Supplement
500 mbar
±0.075 % of URL/
70 bar
±0.14 % of URL/
70 bar
40 bar
±0.075 % of URL/70 bar
±0.14 % of URL/70 bar
500 mbar
±0.14 % of URL/
70 bar
±0.14 % of URL/
70 bar
40 bar
±0.14 % of URL/70 bar
±0.14 % of URL/70 bar
= 70 bar).