Depending on the failure reason and measures taken, the steps
described in chapter "Set up" must be carried out again, if necessary.

9.3 Exchanging the electronics module

In case of a defect, the electronics module can be exchanged by the
user against an identical type. If no electronics module is available on
side, the module can be ordered from the agency serving you.
In Ex applications only one instrument and one electronics module with
respective Ex approval may be used.
The sensor serial number is required to order an exchange electronics.
You will find this number on the type label of the instrument or on the
delivery note.
9.4 Exchanging the electronics module
In case of a defect, the electronics module can be exchanged by the
user for an identical type. If no electronics module is available on site,
the module can be ordered from the agency serving you.
Ordering and exchange are possible with or without sensor serial
number. The electronics module with serial number includes order-
specific data such as factory setting, seal material etc. These are not
included in the electronics module without serial number.
The serial number is stated on the type label of VEGADIF 65 or on the
delivery note.

9.5 Software update

The software version of VEGADIF 65 can be determined as follows:
on the type label of the electronics
l
via the indicating and adjustment module
l
via PACTware
l
You can view all software histories on our website www.vega.com.
Make use of this advantage and get registered for update information
via e-mail.
The following components are required to update the sensor software:
Sensor
l
Voltage supply
l
VEGACONNECT
l
PC with PACTware
l
Current sensor software as file
l
At "www.vega.com/downloads" go to "Software". Select under "plics
instruments and sensors" the suitable instrument series. Load the zip
file via the right mouse key with "Save target as" e.g. on the desktop of
your PC. Extract all files available in the zip file, e.g. to the desktop.
VEGADIF 65 • Foundation Fieldbus

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents