Front Panel

Most of the front panel controls duplicate functionality available through the
touch screen display. They are covered in more detail in the Basics section
and in the HDO8000A Oscilloscopes Operator's Manual. Below are some special
front panel controls.
A
Auto Setup turns on and configures all channels with a signal
attached. The edge trigger level is set to the signal mean.
B
Default Setup, on the other hand, returns all channels to the factory
default settings.
The Print button captures the screen and handles it according to your
C
Print setting (print to file on USB drive, create LabNotebook entry, etc.).
It can be set to capture only the grid rather than the full screen.
D
The Touch Screen button enables or disables touch screen
functionality.
E
Clear Sweeps resets math and measurement counters.
F
The front panel Adjust knob changes the value in a highlighted data
entry field when turned. Pushing the Adjust knob toggles between
coarse (large increment) or fine (small increment) adjustments when
the knob is turned.
All the knobs on the front panel function one way if
turned and another if pushed like a button. The first
label describes the knob's principal "turn" action,
while the second describes its "push" action.
Many front panel buttons light to indicate the
function is active.
USER INTERFACE
C
D
A
B
F
26
E