DNA 7500 and DNA 12000 Sample Well Results

To review the results of a specific sample, select the sample name in the
tree view and highlight the Results sub-tab. The electropherogram of the
sample well window should resemble the one shown here.
Figure 4
Major features for a successful DNA sample run are:
• All sample peaks appear between the lower and upper marker peaks. If
• Flat baseline
• Baseline readings at least 5 fluorescence units (see Zero Baseline in the
• Marker readings at least 3 fluorescence units higher than baseline readings.
• Both marker peaks well resolved from sample peaks (depends on sample).
Checking Your Agilent DNA 7500 and DNA 12000 Assay Results
DNA peaks of a successful sample run
some sample peaks are outside the marker bracket, adjust the upper or
lower marker. Please refer to the 2100 Expert User's Guide or Online Help
for details.
User's guide or Online Help for details of how to see the baseline readings).
DNA 7500 and DNA 12000 Sample Well Results